Figure 8-14: HEATER LED on Membrane Keypad
Potential cause
The environment where the transmitter is installed exceeds the ambient temperature
requirements, or the heat due to convection is causing case temperature to rise above the
limit.
Recommended actions
1. Place a spool piece between the stack flange and the transmitter flange.
2. Relocate the transmitter to a cooler area.
Fault 7, LOI
When Fault 7 is detected, the LOI displays the
Board Temp Hi
message.
Reference Manual
Troubleshoot
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
131