Figure 1-11: Rosemount SPS 4001B Single Probe Autocalibration Option (with
Reference Air Option) Installation with Remote Electronics
A. Duct
B. Calibration gas
C. Reference air
D. Rosemount SPS 4001B
E. Line voltage
F. Logic I/O
G. 4 to 20 mA signal
H. Line voltage
I. Remote electronics
J. Rosemount Oxymitter 4000
K. Adapter plate
L. Stack
M. Gases
A source of instrument air is optional at the transmitter for reference air use. Since the unit
can be equipped with an in-place calibration feature, you can make provisions to
permanently connect calibration gas bottles to the transmitter. If you are permanently
connecting the calibration gas bottles, use a check valve next to the calibration fittings on
the integral electronics. The check valve is to prevent breathing of the calibration gas and
subsequent flue gas condensation and corrosion. The check valve is in addition to the stop
valve in the calibration gas kit and solenoid valves in the Rosemount IMPS 4000 or SPS
4001B.
Note
The integral electronics is rated NEMA 4X (IP66) and is capable of operation at
temperatures up to 185 °F (85 °C). The optional LOI is also rated for operation at
temperatures up to 185 °F (85 °C). The infrared keypad functionality will degrate at
temperatures above 158 °F (70 °C).
Retain the transmitter's original packaging in case you need to ship the components to
another site. This packaging is designed to protect the product.
Reference Manual
Description and specifications
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
29
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...