Line voltage selection
Jumper (install)
240 Vac
JM6, JM5, JM1
Probe heater voltage selection
Jumper (install)
World Class probe (44 V)
JM10
218 probe (115 V)
JM9
Rosemount World Class "Direct Replacement"
Probe (115 V) or Rosemount Oxymitter Direct
Replacement Probe (115 V)
JM9
CAUTION
If you select incorrect heater voltage, damage to the transmitter may occur.
Always update the relevant labeling to reflect the set voltage.
Always disconnect the voltage from heater power supply and analog electronics (if used)
before changing jumpers.
Line voltage selection
Jumper (install)
100/120 Vac
JM4, JM1
220/240 Vac
JM5
(1)
100 Vac operation requires transformer part number 1M02961G02.
Heater power
Jumper
Remote
Remove JM2
On
Install JM2
Probe heater voltage selection
Jumper (install)
Rosemount World Class probe (44 V)
JM7
218 probe (115 V)
JM8
Rosemount Direct Replacement World Class or
DR Oxymitter
JM8
Electronics selection
Jumper
Analog (existing)
Install JM3, JM6
Digital (next generation)
Remove JM3, JM6
Reference Manual
Install
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
61
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...