Figure 2-9: Probe Mounting for Metal Wall Stack or Duct Construction
A. Minimum diameter hole in wall.
B. Stack or duct metal wall.
C. Weld or bolt mounting plate to metal wall of stack or duct. Joint must be airtight.
Figure 2-10: Probe Mounting for Masonry Wall Stack Construction
A. Field weld pipe to mounting plate.
B. Pipe 88,9 mm (3.5 in.) sched 40 pipe sleeve (not by Emerson) length by customer
C. Masonry stack wall
D. Outside wall surface
E. Joint must be airtight.
F. Bolt mounting plate to outside wall surface.
Reference Manual
Install
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
47
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...