1. Blind cover
12. Screw
24. Junction box
1A. Window cover
13. Lock washer
25. Adapter ring
2. Electronic assembly
14. Cable clamp
26. O-ring
3. Screw
15. Terminal block
27. Harness assembly
4. Membrane keypad
16. Captive screw
28. Spacer block
4A. LOI module
17. Blind cover
29. Connecting cable, signal
5. Microprocessor board
17A. O-ring
30. Connecting cable, heater
6. Analog board
18. Screw
31. Interconnecting cable
7. Fuse cap
19. Cover lock
32. Terminal block, remote
probe head
8. Fuse
20. Captive washer
33. Probe
9. Power supply board
21. Lock washer
34. Housing
10. Captive screw
22. Screw
11. Housing
23. Mounting kit
Note
The electronic assembly (Item 2) consists of items 3 through 10.
Note
3. Connect 4-20 mA signal and calibration handshake/logic/I/O leads.
2.3.2
4-20 mA signal
The 4-20 mA signal represents the O
2
value and can also operate the Rosemount
™
751
Loop LCD Display or any other loop powered display. Superimposed on the 4-20 mA signal
is HART
®
information that is accessible through a handhedl communicator or AMS
software.
2.3.3
Rosemount
™
World Class Intelligent Field Transmitter
setup
The DR probe operates with a 115 Vac heater. Ensure that the voltage selection jumpers in
the IFT or HPS, if used, are set properly.
Always disconnect line voltage from intelligent field transmitters before changing
jumpers.
Line voltage selection
Jumper (install)
100 Vac
JM3, JM7, JM2
120 Vac
JM8, JM7, JM1
220 Vac
JM6, JM5, JM2
Install
Reference Manual
October 2019
00809-0100-4340
60
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...