
9.3.8
Replace heater strut
Do not attempt to replace the heater strut until you have eliminated all other possibilities
for poor performance.
WARNING
The transmitter, probe tube, and flame arrestor hub can be hot (up to 572 °F [300 °C]) in
the region of the probe heater. This can cause severe burns.
Wear heat resistant gloves and clothing when removing probe.
Do not attempt to work on the probe until it has cooled to room temperature.
If you need to replace the heater strut, order a replacement heater strut. See
Procedure
to remove the transmitter from the stack
or duct.
2. Remove oxygen sensing cell per
.
CAUTION
Damage to the aluminum housing can occur.
Do not force the probe housing when installing or removing it from the integral
electrical barrier/feed-through (
).
) and washers (21) from the probe tube
assembly (23). Remove the probe tube assembly from the housing (11).
Once you have separated the probe and housing, spring tension releases, and the
heater strut assembly (32) moves up.
4. Remove strut pressure clamp (35).
5. Disconnect the heater and signal wire connectors from the mating connectors on
the heater strut assembly (32).
6. Remove tube clamps (33). Carefully pull the CAL and REF gas silicone tubes (34)
from the CAL and REF gas ports.
7. Remove gas port fittings (29) from the CAL, REF, and VENT ports.
8. Grasp the wire loop and carefully slide the heater strut assembly (32) out of the
probe tube.
See
.
Maintenance and service
Reference Manual
October 2019
00809-0100-4340
164
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...