-40 °C ≤ Ta ≤ +70 °C
Probe assemblies without the frame arrestor end: "-NF"
II 2/- G
Ex d IIB+H2 T4 Gb/-
-40 °C ≤ Ta ≤ +70 °C
Remote electronics: "Split architecture":
II 2 G
Ex d e IIB+H2 T5 Gb
-40 °C ≤ Ta ≤ +70 °C
Special conditions for safe use (X):
1. Mounting flange temperature shall not exceed 115 °C.
2. Non-flame arrestor probe versions "NF" must have the probe tube mounted in a Safe
Area.
3. When the probe tube is mounted in Explosive Areas using flame arrestor end to
complete the assembly, calibration lines that travel in and out of the equipment
bringing reference gas, must not contain a pressure higher than 1.1 times the
atmospheric pressure; these calibration lines shall not contain pure oxygen,
acetylene, or combustible gases other than the gases for which this application has
been investigated for: Group IIB + H2 gases.
4. There are no interchangeable enclosure components.
5. Contact the original manufacturer for information of flameproof joint dimensions.
6. Fastener property class must be A2-70.
E.5
Rosemount
™
Xi, X-Stream Xi Remote Interface,
for general locations certification
E.5.1
USA/Canada
CSA Certificate
1913435
Standards
CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1:2004 (2nd Ed.), CAN/
CSA-E60079-1:07, CAN/CSA-C22.2 No. 94-M91 (R2001), CAN/CSA
C22.2 No. 60529:05, ANSI/ISA-12.00.01-2005 (IEC 60079-0 Mod),
ANSI/ISA-12.22.01-2005 (IEC 60079-1 Mod), UL 50 (11th Ed.), IEC
60529 (Edition 2.1-2001-02)
Markings
FM Certificate
FM17NUS0007
Standards
FM Class 3810:2005, ANSI/NEMA 250:1991
Markings
Type 4X
Reference Manual
Rosemount
™
Oxymitter
™
product certifications
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
191
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...