a. 4-20 mA signal: The 4-20 mA signal represents the O
2
value and can also
operate the Rosemount 751 LCD Loop Display or any other loop-powered
display. Superimposed on the 4-20 mA signal is HART
®
information that is
accessible through a handheld communicator or AMS software.
WARNING
If using a Rosemount IMPS 4000 or SPS 4001B, install it in a non-hazardous,
explosive-free environment.
b. Calibration handshake/Logic/I/O: You can use this output signal to trigger an
alarm or provide a calibration handshake signal to a Rosemount IMPS 4000 or
SPS 4001B.
If not using autocalibration, a common bi-directional logic contact is
provided for any of the equipment alarms listed in
Table 2-8: Diagnostic LEDs
LED
Flashes
Status
4-20 mA line
Fault
Recoverable
Heater C/T
1
Open
Dependent on position 3 of
SW2
1
No
2
Shortened
Dependent on position 3 of
SW2
2
No
3
Reversed
Dependent on position 3 of
SW2
3
No
4
A/D Comm Error
Dependent on position 3 of
SW2
4
No
Heater
1
Open
Dependent on position 3 of
SW2
5
No
2
High High Temp
Dependent on position 3 of
SW2
6
No
3
High Case Temp
Dependent on position 3 of
SW2
7
Yes
4
Low Temp
Dependent on position 3 of
SW2
8
Yes
5
High Temp
Dependent on position 3 of
SW2
9
Yes
O
2
cell
1
High mV
Dependent on position 3 of
SW2
10
Yes
3
Bad
Track O
2
11
Yes
4
EEprom Corrupt
Dependent on position 3 of
SW2
12
No
Calibration
1
Invalid Slope
Track O
2
13
Yes
2
Invalid Constant
Track O
2
14
Yes
3
Last Calibration Failed
Track O
2
15
Yes
Install
Reference Manual
October 2019
00809-0100-4340
56
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...