Figure 3-1: Electronics Housing Terminals and Membrane Keypad
A. Rosemount
™
Hazardous Area Oxymitter 4000 electronics housing
B. Terminal block
C. Ground lugs
D. Logic I/O
E. 4-20 mA signal
3. Install the cover over terminal block and secure with captive washer, cover lock, and
screw.
3.4
Configuration
Located on the microprocessor board, the top board, are two switches that configure
outputs for the transmitter.
WARNING
The HART
®
option is not protected by energy limiting barriers.
Do not interface with the HART option from within the hazardous area.
Route the 4-20 mA cables and make the connections outside the hazardous area.
This is the case even when using the intrinsically safe version of the handheld
communicator.
Configure transmitter with membrane keypad
Reference Manual
October 2019
00809-0100-4340
70
Emerson.com/Rosemount