Figure 1-16: Rosemount Oxymitter Electronics and Rosemount Xi Maximum Loop
Resistance is Determined by the Voltage Level of the External Power Supply as
Described by
A. Load (ohms)
B. Minimum lift off
C. Power supply voltage
D. Maximum
E. Without HART
®
communicator
F. Operating region
1.11.3
Physical specifications
Materials of
construction
Transmitter probe: Process-wetted materials are 316L or 304 stainless
steel.
Rosemount
™
Oxymitter Transmitter electronics housing (integral to
probe or remote mounted): Low copper aluminum IP66 (NEMA 4X)
with reference air exhaust port piped to clean area
Optional Rosemount Xi Electronics: NEMA 4X, polycarbonite material
Process
connections
2-in. 150# (4.75-in. [121 mm] bolt circle); DIN (5.71-in. [145 mm] bolt
circle)
Note
Flanges are flat-faced and for mounting only. Flanges are not
pressure-rated. Spool piece part numbers are available to offset
electronics housing from hot ductwork. Many adapter flanges are
available to mate to existing flanges.
Orientation
Vertical or horizontal mount
Reference Manual
Description and specifications
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
35
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...