2.3.1
Connect line voltage
Procedure
1. Remove screw, cover lock, captive washer, and left side blind cover from the
remote electronics.
2. Connect the line (or L1) wire to the L1 terminal and the neutral (or L2) wire to the N
terminal.
The transmitter will automatically configure itself for 90 - 250 Vac line voltage and
50/60 Hz.
WARNING
Electrical shock
Install the power terminal cover.
Figure 2-18: Electrical Installation: Rosemount
™
Hazardous Area Oxymitter
4000 with Remote Electronics
Reference Manual
Install
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
59