Figure 8-10: HEATER LED on Membrane Keypad
Recommended actions
1. Remove power.
2. Remove the electronic assembly per .
3. Using a multimeter, measure the resistance across the terminals of heater
connector, J8.
The measurement should be approximately 72 ohms. If the heater is open, see
Fault 5, LOI
When Fault 5 is detected, the LOI displays the
O2 Heater Open
message.