WARNING
Failure to install covers and ground leads could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installation.
WARNING
To meet the safety requirements of IEC 61010 (EC requirement) and ensure safe
operation of this equipment, connect to the main power supply through a circuit
breaker (minimum 10 A) which will disconnect all current-carrying conductors during a
fault situation.
Make sure this circuit breaker includes a mechanically operated isolating switch.
If there is no switch, locate another external means of disconnecting the power supply
close by.
Circuit breakers or switches must comply with a recognized standard, such as IEC
60947.
WARNING
The instrument is heavy.
Use proper lifting and carrying procedures to avoid personal injury.
WARNING
To maintain explosion-proof protection, all cable entry devices and blanking elements for
unused apertures must be certified flameproof and suitable for the conditions of use.
Make sure they are properly installed.
Note
To maintain CE compliance, ensure a good connection exists between the mounting
flange bolt and earth.
2.2.1
Connect line voltage
Procedure
1. Remove screw, cover lock, and captive washer. Remove cover from terminal block.
Install
Reference Manual
October 2019
00809-0100-4340
52
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...