Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose
Device Tools
>
Configuration
>
Process Measurement
>
Concentration Measurement
.
2. Scroll to Step 2, set
Active Matrix
to the matrix you want to use and click
Change Matrix
.
5.2.3 Set up concentration measurement using the Field
Communicator
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Enable the concentration measurement application using the Field
Communicator
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
•
The concentration measurement application must be licensed on your transmitter.
•
The concentration measurement application cannot be enabled at the same time as the API referral
application or the Advanced Phase Measurement application. They must be disabled first.
Procedure
1. Choose
Overview
>
Device Information
>
Licenses
>
Enable/Disable Applications
and ensure that
Volume Flow Type
is set to Liquid.
2. Choose
Overview
>
Device Information
>
Licenses
>
Enable/Disable Applications
.
3. Enable the concentration measurement application.
Set reference temperature values for specific gravity using the Field
Communicator
When
Derived Variable
is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
To check the setting of
Derived Variable
, choose
Configure
>
Manual Setup
>
Measurements
>
Optional
Setup
>
Conc Measurement
>
CM Configuration
.
Important
Do not change the setting of
Derived Variable
. If you change the setting of
Derived Variable
, all existing
concentration matrices will be deleted from transmitter memory.
Configure process measurement applications
Configuration and Use Manual
April 2019
MMI-20048166
64