Configure Volume Flow Measurement Unit for liquid applications
Configure Volume Flow Cutoff
4.3.1 Configure Volume Flow Type for liquid applications
Display
Menu
>
Configuration
>
Process Measurement
>
Flow Variables
>
Volume Flow Settings
>
Flow Type
>
Liquid
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
>
Volume Flow Type
>
Liquid Volume
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Optional Setup
>
GSV
>
Volume Flow Type
>
Liquid Volume
Volume Flow Type
controls whether liquid or gas standard volume flow measurement will be used.
Restriction
If you are using the API referral application, you must set
Volume Flow Type
to Liquid. Gas standard volume
measurement is incompatible with the API referral application.
Restriction
If you are using the concentration measurement application, you must set
Volume Flow Type
to Liquid.
Gas standard volume measurement is incompatible with the following applications:
•
Concentration measurement
Procedure
Set
Volume Flow Type
to Liquid.
4.3.2 Configure Volume Flow Measurement Unit for liquid
applications
Display
Menu
>
Configuration
>
Process Measurement
>
Flow Variables
>
Volume Flow Settings
>
Units
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
>
Volume Flow Rate Unit
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Flow
>
Volume Flow Unit
Volume Flow Measurement Unit
specifies the unit of measurement that will be displayed for the volume
flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure
Volume Flow Measurement Unit
, be sure that
Volume Flow Type
is set to Liquid.
Procedure
Set
Volume Flow Measurement Unit
to the unit you want to use.
•
Default: l/sec (liters per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Configuration and Use Manual
Configure process measurement
MMI-20048166
April 2019