c) Read the mA current at the receiving device and compare it to the transmitter output.
The readings do not need to match exactly. If the values are slightly different, you can correct
the discrepancy by trimming the output.
d) Press
OK
.
e) Select
20 mA
.
f) Read the mA current at the receiving device and compare it to the transmitter output.
The readings do not need to match exactly. If the values are slightly different, you can correct
the discrepancy by trimming the output.
g) Press
OK
.
h) Choose
End
.
2. Test the Frequency Output(s).
a) Choose
Service Tools
>
Simulate
>
Simulate Outputs
and select the frequency output to test.
b) Select the frequency output level.
c) Press
OK
.
d) Choose
End
.
3. Test the Discrete Output(s).
a) Choose
Service Tools
>
Simulate
>
Simulate Outputs
and select the discrete output to test.
b) Choose
Off
.
c) Verify the signal at the receiving device.
d) Press
OK
.
e) Choose
On
.
f) Verify the signal at the receiving device.
g) Press
OK
.
h) Choose
End
.
Postrequisites
•
If the mA Output readings are within 20 microamps of the expected values, you can correct this
discrepancy by trimming the output.
•
If the discrepancy between the mA Output readings is greater than 20 microamps, or if at any step the
reading was faulty, verify the wiring between the transmitter and the remote device, and try again.
Troubleshooting
Configuration and Use Manual
April 2019
MMI-20048166
192