15.2.36 [116] Temperature Overrange (API referral)
Alert
Process Aberration
Cause
The line temperature is outside the range of the API table.
Recommended actions
1. Ensure that process temperature is within the range of the API table.
2. Verify the configuration of the API referral application and related parameters.
15.2.37 [117] Density Overrange (API referral)
Alert
Process Aberration
Cause
The line density is outside the range of the API table.
Recommended actions
1. Ensure that process density is within the range of the API table.
2. Verify the configuration of the API referral application and related parameters.
15.2.38 [118] Discrete Output 1 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant state. A loop test may be in progress.
Recommended actions
Check to see if a loop test is in process (the output is fixed).
15.2.39 [120] Curve Fit Failure (Concentration)
Alert
Configuration Error
Cause
The transmitter was unable to calculate a valid concentration matrix from the current data.
Recommended actions
1. Verify the configuration of the concentration measurement application.
Troubleshooting
Configuration and Use Manual
April 2019
MMI-20048166
170
Micro Motion 4200 Transmitters