Table 12-1: SMV abort codes
(continued)
Code
Description
Recommended actions
10
State in progress
NOT AVAILABLE
11
Verification Complete
NOT AVAILABLE
12
Transmitter in fault and SMV
cannot run
View the Alerts present on the device and take any necessary
actions to clear them before proceeding.
13
No factory air verification
Perform factory calibration on air.
14
No field reference point
Perform a field reference point procedure.
15
Drive frequency drift from carrier
frequency
Ensure temperature, flow, and density are stable. Start again.
Other
General abort
Repeat the test. If the test aborts again, contact customer
service.
12.1.3 Set up SMV automatic execution
You can set up and run a single test at a user-defined future time. You can also set up and run tests
automatically on a regular schedule.
Automatic execution of SMV is managed from the transmitter. You do not need a connection from an
external configuration tool.
Tip
The time between test runs must be between 1 and 1000 hours. The time to the first test run can be any
positive floating number.
Important
Only the 20 most recent SMV results are stored. For SMV Basic versions, the last automatic execution result
can be viewed, while all 20 results can be viewed with SMV licensed. To view or chart these results using an
external tool, you must upload them from the transmitter.
Related information
Set up SMV automatic execution using the display
Set up SMV automatic execution using ProLink III
Set up SMV automatic execution using the Field Communicator
Set up SMV automatic execution using the display
Procedure
1. Choose
Menu
>
Operations
>
Smart Meter Verification
>
Schedule Verification
.
2. To schedule a single test:
a) Set
Hours to 1st Run
to the number of hours to elapse before the test is run.
b) Set
Hours Between
to
0
.
3. To schedule a recurring execution:
a) Set
Hours to 1st Run
to the number of hours to elapse before the first test is run.
Measurement support
Configuration and Use Manual
April 2019
MMI-20048166
130