Table 7-2: Options for alerts and conditions
(continued)
Alert
Conditions
Name
Description
Ignorable
•
Severity configurable: No
•
Fault Timeout applicable: Yes
[016] Sensor Temperature
(RTD) Failure
The value computed for the
resistance of the line RTD is outside
limits.
No
[017] Sensor Case
Temperature (RTD) Failure
The values computed for the
resistance of the meter and case
RTDs are outside limits.
No
Configuration Error
•
Default severity: Failure
•
Severity configurable: No
•
Fault Timeout applicable: Yes
[020] Calibration Factors
Missing
Some calibration factors have not
been entered or are incorrect.
No
[021] Incorrect Sensor Type
Transmitter verification of sensor
circuits and characterization has
produced a discrepancy. The
transmitter cannot operate the
sensor.
Yes
[030] Incorrect Board Type
The firmware or configuration
loaded in the transmitter is
incompatible with the board type.
No
Password Not Set
Display security has been enabled
but the display password has not
been changed from the default
value.
No
Time Not Entered
The system time has not been
entered. The system time is required
for diagnostic logs.
Yes
[120] Curve Fit Failure
(Concentration)
The transmitter was unable to
calculate a valid concentration
matrix from the current data.
No
Tube Not Full
•
Default severity: Failure
•
Severity configurable: No
•
Fault Timeout applicable: Yes
[033] Insufficient Pickoff
Signal
The signal from the sensor pickoffs is
insufficient for operation. (Enhanced
core processor only)
Yes
Extreme Primary Purpose Variable
•
Default severity: Failure
•
Severity configurable: No
•
Fault Timeout applicable: Yes
[005] Mass Flow Rate
Overrange
The measured flow is outside the
sensor's flow limits.
No
[008] Density Overrange
The measured density is above
10 g/cm
³
.
No
Transmitter Initializing
•
Default severity: Failure
•
Severity configurable: No
•
Fault Timeout applicable: No
[009] Transmitter Initializing/
Warming Up
The transmitter is in power-up
mode.
No
Function Check in Progress
[104] Calibration in Progress
A calibration is running.
No