2. Test the Frequency Output(s).
a) Choose
Menu
>
Service Tools
>
Output Simulation
and select the frequency output to test.
b) Set
Simulation Value
to
1
.
c) Start the simulation.
d) Read the frequency signal at the receiving device and compare it to the transmitter output.
e) Choose
New Value
.
f) Set
Simulation Value
to
1450
.
g) Start the simulation.
h) Read the frequency signal at the receiving device and compare it to the transmitter output.
i) Choose
Exit
.
3. Test the Discrete Output(s).
a) Choose
Menu
>
Service Tools
>
Output Simulation
and select the discrete output to test.
b) Set
Simulation Value
to ON.
c) Start the simulation.
d) Verify the signal at the receiving device.
e) Choose
New Value
.
f) Set
Simulation Value
to OFF.
g) Start the simulation.
h) Verify the signal at the receiving device.
i) Choose
Exit
.
Postrequisites
•
If the mA Output readings are within 20 microamps of the expected values, you can correct this
discrepancy by trimming the output.
•
If the discrepancy between the mA Output readings is greater than 20 microamps, or if at any step the
reading was faulty, verify the wiring between the transmitter and the remote device, and try again.
15.15.2 Perform loop tests using ProLink III
Procedure
1. Test the mA output(s).
a) Choose
Device Tools
>
Diagnostics
>
Testing
and select the mA output to test.
b) Enter
4
in
Fix to:
.
c) Click
Fix mA
.
d) Read the mA current at the receiving device and compare it to the transmitter output.
Troubleshooting
Configuration and Use Manual
April 2019
MMI-20048166
190
Micro Motion 4200 Transmitters