2. Verify that the transmitter is configured correctly for the connected sensor.
15.2.29 [106] Burst Mode Enabled AI or AO Simulate Active
Alert
Burst Mode
Cause
The device is in HART burst mode.
Recommended actions
1. No action required.
2. If desire, reconfigure the alarm severity to ignore.
15.2.30 [107] Power Reset Occurred
Alert
Power
Cause
The transmitter has been restarted.
Recommended actions
1. No action required.
2. If desire, reconfigure the alarm severity to ignore.
15.2.31 [110] Frequency Output
X
Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the scaling of the Frequency Output.
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
15.2.32 [111] Frequency Output 1 Fixed
Alert
Output Fixed
Troubleshooting