Variable
is set to Mass Flow Rate, the mass flow rate reported via the mA Output is controlled by the higher of
the two cutoff values.
Mass Flow Cutoff
affects all reported values and values used in other transmitter behavior (e.g., events
defined on mass flow).
mA Output Cutoff
affects only mass flow values reported via the mA Output.
Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
•
mA Output Process Variable
: Mass Flow Rate
•
Frequency Output Process Variable
: Mass Flow Rate
•
mA Output Cutoff
:
10
g/sec
•
Mass Flow Cutoff
:
15
g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all
internal processing.
Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
•
mA Output Process Variable
: Mass Flow Rate
•
Frequency Output Process Variable
: Mass Flow Rate
•
mA Output Cutoff
:
15
g/sec
•
Mass Flow Cutoff
:
10
g/sec
Result:
•
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
—
The mA Output will report zero flow.
—
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
•
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.3 Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported. The
volume total and volume inventory are derived from volume flow data.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one
or the other.
Related information
Configure Volume Flow Type for liquid applications
Configure process measurement
Configuration and Use Manual
April 2019
MMI-20048166
28
Micro Motion 4200 Transmitters