7.2.4 Configure Fault Timeout
Display
Menu
>
Configuration
>
Alert Setup
>
Output Fault Actions
>
Fault Timeout (sec)
ProLink III
Device Tools
>
Configuration
>
Fault Processing
>
Fault Timeout
Field Communicator
Configure
>
Alert Setup
>
Output Fault Actions
>
General
>
Fault Timeout
Fault Timeout
controls the delay before fault actions are performed.
The fault timeout period begins when the transmitter detects an alert condition.
•
During the fault timeout period, the transmitter continues to report its last valid measurements.
•
If the fault timeout period expires while the alert is still active, the fault actions are performed.
•
If the alert condition clears before the fault timeout expires, no fault actions are performed.
Restriction
•
Fault Timeout
is not applied to all alerts. For some alerts, fault actions are performed as soon as the alert
condition is detected. See the list of alerts and conditions for details.
•
Fault Timeout
is applicable only when
Alert Severity
= Failure. For all other settings of
Alert Severity
,
Fault Timeout
is irrelevant.
Procedure
Set
Fault Timeout
as desired.
•
Default: 0 seconds
•
Range: 0 to 60 seconds
If you set
Fault Timeout
to 0, fault actions are performed as soon as the alert condition is detected.
7.2.5 Alerts, conditions, and configuration options
Table 7-2: Options for alerts and conditions
Alert
Conditions
Name
Description
Ignorable
Electronics Failed
•
Default severity: Failure
•
Severity configurable: No
•
Fault Timeout applicable: No
[018] EEPROM Error
(Transmitter)
There is an internal memory
problem with the transmitter.
No
[019] RAM Error (Transmitter) There is a ROM checksum mismatch
in the transmitter or the RAM
address location cannot be written
in the transmitter.
No
Watchdog Error
The watchdog timer has expired.
No
Verification of mA Output 1
Failed
The reading of the mA input does
not match the reading of mA Output
1.
No
Sensor Failed
•
Default severity: Failure
[003] Sensor Failed
The pickoff amplitude is too low.
No
Configuration and Use Manual
Configure device options and preferences
MMI-20048166
April 2019