If your sensor does not show an
FD
value, contact customer service.
If your sensor tag does not show a
DT
or
TC
value, enter the last 4 characters of the density calibration factor.
In the sample tag shown above, the value is shown as 4.44.
2.9 Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available
method.
Procedure
•
Read the value for
Mass Flow Rate
on the transmitter display, which is the default initial display.
•
Connect to the transmitter with ProLink III and read the value for
Mass Flow Rate
in the
Process Variables
panel.
•
Connect to the transmitter with the Field Communicator and read the value for
Mass Flow Rate
.
Online
>
Overview
>
Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
•
Check the characterization parameters.
•
Review the troubleshooting suggestions for flow measurement issues.
2.10 Verify the zero
Display
Menu
>
Service Tools
>
Verification & Calibration
>
Meter Zero
>
Zero Verification
ProLink III
Device Tools
>
Calibration
>
Smart Zero Verification and Calibration
>
Verify Zero
Field Communicator
Service Tools
>
Maintenance
>
Calibration
>
Zero Calibration
>
Perform Zero Verify
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field
zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the
following is true:
•
The zero is required by site procedures.
•
The stored zero value fails the zero verification procedure.
Prerequisites
Important
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the
zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Procedure
1. Prepare the meter:
Configuration and Use Manual
Quick start
MMI-20048166
April 2019
Configuration and Use Manual
17