Calibration tables may be copied from each 6560 Processor Module onto an online server,
and those can be downloaded into the same 6560 Processor Module. Emerson Online
Product Support personnel and Online Systems Engineers can assist customers with this
type of operation.
Calibration circuit inputs and outputs are shown in
B.1.1
Calibration circuits input and output
Calibration circuits input and output
Table B-1:
Calibration circuit
Input
Output
TSG (Test Signal Generator)
(external) signal
TSG table (internal) TSG signal
GS (Gross Scan)
TSG signal
GS table
DSP (Digital Signal Processor)
TSG signal
DSP table
Transient
TSG signal
Transient table
B.1.2
Test Signal Generator (TSG) calibration
All CPU boards provided by Emerson are shipped with a calibrated TSG circuit. If the CPU
board is purchased as part of a system, the entire system is calibrated using the CPU board
TSG circuit.
The TSG circuit is the only element that requires an external piece of test equipment and
special connection cables. TSG calibration requires:
•
Digital Multi Meter (such as the HP 34401A)
•
computer configured as online server
•
special calibration utility program (DHM)
•
cable that connects the computer to the 6560 Processor module Ethernet port
•
cable that connects the computer to the Digital Multi Meter
•
cable that connects the Digital Multi Meter to the test port on the unit
TSG calibration is unique in that it requires a separate test instrument, unique cables, and a
computer that is configured as an online server. It is recommended that qualified Emerson
Online Systems Engineers perform TSG calibration or recalibration.
B.1.3
Gross Scan (GS) calibration
GS calibration:
•
uses a 6560 Processor module's TSG output signal.
•
does not require that any wire harnesses be disconnected.
•
is completed in about 10 minutes.
•
does not require any special cables or test equipment.
System calibration
84
MHM-97453, Rev 1
Summary of Contents for Machinery Health Expert CSI 2600
Page 1: ...User Guide MHM 97453 Rev 1 April 2014 CSI 2600 Machinery Health Expert User Guide ...
Page 8: ...CSI 2600 back view Figure 2 2 Introduction to the CSI 2600 4 MHM 97453 Rev 1 ...
Page 12: ...Introduction to the CSI 2600 8 MHM 97453 Rev 1 ...
Page 26: ...Getting started 22 MHM 97453 Rev 1 ...
Page 64: ...Software configuration 60 MHM 97453 Rev 1 ...
Page 96: ...Data types 92 MHM 97453 Rev 1 ...
Page 100: ...D 2 Internal wiring diagram for the CSI 2600 ...
Page 106: ...Troubleshooting 102 MHM 97453 Rev 1 ...
Page 109: ...Index MHM 97453 Rev 1 105 ...