6.10
Change databases when moving the unit to a
new machine
Because the CSI 2600 is a portable system, analysts must ensure that measurements from
one machine are not stored in a database or archive folder for a different one.
CAUTION!
Ensure that the following sequence is observed whenever moving the CSI 2600 from one
monitoring rack or machine to another. Data from one machine could be stored in the
database of another machine if this sequence is performed incorrectly.
CAUTION!
Changing databases will reinitialize the Transient HDD, which eliminates all stored data.
Extract any data before changing databases.
Prerequisites
Before disconnecting from the first machine, log in to AMS Machinery Manager.
From the
Tools
tab, click
Setup/Communication
and open RBM Network Administration.
Procedure
1.
In the bottom center panel labeled Online Servers, double-click the server.
The Online Server Setup dialog opens.
2.
Click
Stop Data Collection
.
3.
Disconnect the unit from the first machine.
4.
Connect the unit to the second machine.
5.
In Online Server Setup, click the
Edit
button next to Machinery Health Manager
Database—not next to Active Units.
6.
Select the second machine's database when prompted.
7.
In Online Server Setup, click
Start Data Collection.
The system will now store any measurements or archives in the second database.
Data collection and analysis
72
MHM-97453, Rev 1
Summary of Contents for Machinery Health Expert CSI 2600
Page 1: ...User Guide MHM 97453 Rev 1 April 2014 CSI 2600 Machinery Health Expert User Guide ...
Page 8: ...CSI 2600 back view Figure 2 2 Introduction to the CSI 2600 4 MHM 97453 Rev 1 ...
Page 12: ...Introduction to the CSI 2600 8 MHM 97453 Rev 1 ...
Page 26: ...Getting started 22 MHM 97453 Rev 1 ...
Page 64: ...Software configuration 60 MHM 97453 Rev 1 ...
Page 96: ...Data types 92 MHM 97453 Rev 1 ...
Page 100: ...D 2 Internal wiring diagram for the CSI 2600 ...
Page 106: ...Troubleshooting 102 MHM 97453 Rev 1 ...
Page 109: ...Index MHM 97453 Rev 1 105 ...