D.3
Terminal descriptors
Each channel has five terminals. The first two are for the plus (+) and minus (-) signal
inputs. If the associated DIP switch is set to ON, these terminals will also 24 V
constant current accelerometer power.
The second two are for the -24 V power supply for eddy current probes. These terminals
only supply power if an external -24 V power supply is connected to the J19 power input
terminal at the edge of the termination panel.
Note
For the CSI 2600, this connection is not used and the -24 V power is available on the panel-mounted
Phoenix connector on the rear of the case.
The last terminal for each channel is a chassis ground for connecting the sensor cable
shield.