Transient Daughterboard status conditions
(continued)
Table A-6:
LED Color
Status
Priority Comments
Solid Green
OK
4
Normal operation.
System Status LED
The System Status LED indicates the status of the overall system. It indicates the active
status condition with the highest priority of all boards in the system. For example, if the
Test Function generator on the Main Processor board is uncalibrated and the first MSIG
module has a power fault, the LED will show a solid red color to indicate the MSIG module
power fault, which is a “Failure” state.
When all the firmware components are operating as expected, this LED overlays a
“heartbeat” pulse pattern on top of the system status. The heartbeat pattern occurs in a
four-count cycle. The LED is pulsed off briefly during each of the first and second counts,
and then left on during the 3rd and 4th counts. In practice, it gives the appearance of a
human heartbeat. If the heartbeat stops, it indicates a firmware fault has occurred. Many
times the system is capable of recovering from a missed heartbeat. However, if the system
cannot recover quickly, it will automatically reboot itself to clear the fault and will then
resume monitoring.
Server Connect LED
The Server Connect LED indicates when Machinery Health Manager software or the DHM
diagnostic software are connected.
•
A green color indicates that at least one Machinery Health Manager software client
is connected or that the DHM software is connected in the client mode.
•
A red color indicates the DHM software is connected in the Single User mode. In this
state, no other clients can connect.
•
If the LED is off, it indicates that neither of these types of software clients are
connected.
There is no indication of client data transfer, only the presence of at least one established
connection.
Modbus Connect LED
The Modbus Connect LED indicates when a Modbus client, Web Browser, or Transient Live
client are connected.
•
A green color indicates that at least one of these types of clients has established a
connection.
•
If the LED is off, it indicates that none of these types of clients is connected.
The red color is not used with this LED.
There is no indication of client data transfer, only the presence of at least one established
connection.
Specifications
MHM-97453, Rev 1
77
Summary of Contents for Machinery Health Expert CSI 2600
Page 1: ...User Guide MHM 97453 Rev 1 April 2014 CSI 2600 Machinery Health Expert User Guide ...
Page 8: ...CSI 2600 back view Figure 2 2 Introduction to the CSI 2600 4 MHM 97453 Rev 1 ...
Page 12: ...Introduction to the CSI 2600 8 MHM 97453 Rev 1 ...
Page 26: ...Getting started 22 MHM 97453 Rev 1 ...
Page 64: ...Software configuration 60 MHM 97453 Rev 1 ...
Page 96: ...Data types 92 MHM 97453 Rev 1 ...
Page 100: ...D 2 Internal wiring diagram for the CSI 2600 ...
Page 106: ...Troubleshooting 102 MHM 97453 Rev 1 ...
Page 109: ...Index MHM 97453 Rev 1 105 ...