Adjustments
98
Pos : 5. 22 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 T her moc ouple t esti ng @ 9\ mod_1425629857432_9. doc x @ 75276 @ 2 @ 1
4.11 Thermocouple testing
Pos : 5. 23 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 10Adjust ment s_C S-2000/ 1060 M odul T hermoc oupl e t esti ng @ 9\ mod_1425542792773_9. doc x @ 74694 @ @ 1
Fig. 80: Thermocouple testing
Disconnect the plug J1 from the board TH 43 and measure the resistance of the thermocouple
with a digital multimeter, range 100 Ohm.
The resistance must be very low, about 0,5 Ohm.
If the resistance is very high or infinite, then the thermocouple is faulty and must be replaced.
Pos : 6 / 0010 ELTR A/------- Seit enumbruc h ----------- @ 6\ mod_1368796429449_0.doc x @ 41468 @ @ 1
J1
TP1
TP12
TP2
P1
P2
P3
P4
1
J 5
36907
TH43
TP7
TP5
TP4
TP11
TP3
2
1
B
A
2
1
7
8
6
5
4
3
2
1
J3
A B
J 2
TP6
2-15
OHM
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...