Adjustments
76
The outputs of the IR cells (IR base lines) are factory set at 8V when the paths are new and
clean. The peaks come down starting from 8V coming down to a minimum voltage (at the peak
maximum) and they go back to the 8V baseline after completion of combustion. The 8V IR-base
line is obtained through preamplifier adjustment only.
In real operation, with growing contamination resulting in gradually weakened radiation reaching
the detector, the IR cell output base lines will move down to levels below 8V depending on the
grade of contamination. The (negative output) peaks of the cell will also change to smaller
amplitude.
The base line is not regulated any more.
Due to a recalculation formula, the base lines on the screen are always zero volts, the peaks
are positive and they are always of equal amplitude and surface (for same sample and same
weight) regardless the path contamination, regardless the cell base line level, and regardless
the cell output peak amplitude. Consequently the integrated value (of the peak on the screen)
remains the same and of course also the analyses results remain constant at any grade of
contamination.
–
When loading the UNI software select from now on "New cell" mode.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...