Service
108
Tube to squeeze:
Leak location:
13
Exhaust valve (V9)
14
Flow meter (25)
15
CO2-Paths (26)
16
SO3-Trap (27) or oxidizing furnace
tube (28)
17
SO2-Path (29) or flow sensor (30)
18
Regulator valve (V6)
20
Bypass valve (V4), check for 24 V
DC on the coil
21
Purge valve (V3), check for 0 V DC
on the coil
22
Lance valve (V2), check for 24 V DC
on the coil
ATTENTION
do not cause a short circuit while measuring, otherwise the drivers will be
destroyed. See chapter
“
Electronic drivers malfunctioning
”.
–
If the voltages on V2, V3, or V4 are wrong, then replace the electronic drivers.
Pos : 7. 10 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Sol enoi d val ves cleaning ELTR A @ 6\ mod_1369643206541_9.doc x @ 42465 @ 2 @ 1
5.5 Solenoid valves cleaning
Pos : 7. 11 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 15 Ser vic e_C S-2000/1530 Modul Cleani ng s olenoi d val ves @ 9\ mod_1425543480470_9. doc x @ 74766 @ @ 1
Fig. 92: Solenoid valve 2/2
2
4
5
7
1
2
3
4
5
7
1
3
6
6
3 - 1 3 - 1
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...