Faults
40
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Make sure all capacitors are discharged
Fig. 28: Transformer
–
Disconnect the cables (1) from the primary side of the transformer (2) and connect the
cables (1) to a 220 V light bulb (3) which is inserted to a socket. An ordinary desk lamp
can serve for this purpose. In case only 110 V light bulbs are available, connect two of
them in series.
NOTE
Be careful that no short circuit occurs!
After the cables (1) have been connected, plug the analyzer to the mains power again.
–
Set the power switch to pos. 3
–
Enter a weight manually (for example, 500 mg).
–
Press START
–
The word "Analyzing" should appear in the status window.
If not, but "Waiting for stability" appears instead, the reason is that after restarting the
device and the oxygen flow, the infrared base line is drifting. In order to save time, there
is a possibility to start the cycle without waiting for stability.
–
Quit by pressing STOP
–
Go to the "power user" level and click on the button "Stable".
–
The analysis cycle starts then without waiting for base line stability of the cells.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...