Adjustments
70
Pos : 5. 1 / 0025 Ü bersc hrift en/ELTR A 1. Ü berschrif ten/ 1 Adj ust ments eltr a @ 8 \ mod_1396882697255_9. doc x @ 60262 @ 1 @ 1
4
Adjustments
Pos : 5. 2 / 0025 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 G as fl ow c ontroll er adj ust ment and j umper s etti ngs ELTR A @ 6\ mod_1368776355883_9.doc x @ 41371 @ 2 @ 1
4.1 Gas flow controller-adjustment and jumper settings
Pos : 5. 3 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 2000_s er vic e/10Adj ust ments _CS- 2000/1005 Modul G as flow contr oller adj ust ments and j umper s et tings @ 9\ mod_1425542321719_9.doc x @ 74562 @ 344 @ 1
4.1.1 Flow controller for induction furnace operation
Gas flow controller adjustment
Fig. 55: Gas flow controller
–
Set the flow rate with P1 of the HF 42, to about 180 l/h. Observe the flow rate on the
lower flow meter of the analyzer. Wait until the value is stable.
The output of the flow sensor is on TP10 of the HF 42 board; TP15 is GND. With a flow of
180 l/h, the voltage is about 2.5 - 3V.
NOTE
: without flow, the output voltage of the sensor is not 0V but about 0.6V.
–
Adjust with P3 until the voltage on TP7 is 5V.
On TP8, the activity of the pulse width modulator can be checked.
The output voltage to the gas flow regulation valve can be measured on TP3.
With a constant flow of about 180 l/h, it should be between 5 and 15 volts.
–
Check and if necessary readjust the pressure regulator (PR3) in front of the flow
regulating valve.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...