Faults
50
When the furnace is open, the valve ( V2 ) should have 23 V DC on its magnetic coil.
When the furnace is closed, the voltage should be at zero volts.
–
If the voltage on the valve ( V2 ) remains at 23 V DC, no matter whether the furnace is
opened or closed, then there is an electronic problem.
–
If the valve voltage always stays at zero, then either the wiring is broken, see chapter
“
Wiring diagrams
” or the electronic unit is defective.
–
If the voltage on the valve ( V2 ) is correct, then the valve itself is defective, see chapter
“
Cleaning the solenoid valves
”.
Pos : 3. 38 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 C ombusti on wit hout st arting anal ysis ELTR A @ 6 \ mod_1368706920531_9. doc x @ 41209 @ 2 @ 1
3.19 Combustion without starting analysis
Pos : 3. 39 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0610 M odul C ombusti on wit hout st arting anal ysis @ 9\ mod_1425540917709_9. doc x @ 74345 @ @ 1
Induction furnace operation:
In this case, the combustion of the sample begins immediately after the sample enters the
furnace area and before having pressed F5 to start analysis. Should the combustion start just
after closing the furnace then there is a serious fault
Fig. 37: Power switch
The power switch should be set immediately to pos. 1, in order to avoid any overheating of the
generator.
NOTE
Always keep the furnace closed while testing availability of induction power.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...