Discharging the oscillator
11
Attention!
The measuring tip is now held first on the left-hand
and then on the right-hand 4.4 KV connector of the
transformer for approx. 5s
It is necessary to ensure that the tip stays on each
contact for the full 5s.
Attention!
All marked components inside the oscillator are short circuited to ground using the measuring tip.