Faults
66
analysis, the burned sample in the crucible looks like normally combusted sample. If so, check
the bypass valve V4. See chapter
“
Cleaning Solenoid valves
”.
Pos : 3. 68 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 IR signal does not c ome down t o the base line at t he end of anal ysis @ 7\ mod_1381404509510_9. doc x @ 49850 @ 2 @ 1
3.34 IR signal does not come down to the base line at the end of analysis
Pos : 3. 69 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0685 M odul IR signal does not c ome down t o the base line @ 9 \ mod_1425541763229_9. doc x @ 74537 @ 3333 @ 1
There are several different reasons for this effect
1.
Long combustion of a very low carbon sample:
Fig. 51: Programm screenshot (graph example 1)
It appears like a long tailing at the end of the peak. It appears in the low carbon range only and
the distance to the base line is in the range of 10 to 50mV. Depending on the blanks of the
crucibles and on the oxygen purity, the level of the tailing can be as high as 100mV. When the
burning is extremely long, the level of the low carbon tailing even starts to grow.
Possible solutions:
-
Reduce the analysis time
-
Increase the comparator level of the low carbon range,
-
Reduce the generator time.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...