Service
115
Pos : 7. 20 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Gas s aving also f or resis tanc e f urnac e @ 9\ mod_1425632688373_9.doc x @ 75312 @ 2 @ 1
5.10 Gas saving also for resistance furnace
Pos : 7. 21 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 15 Ser vic e_C S-2000/1555 Modul G as s avi ng f or resis tanc e f urnac e @ 9\ mod_1425543689791_9. doc x @ 74826 @ @ 1
If this feature is required, the CS-2000 can be upgraded as follows. But the customer must take
care to have oxygen flow first before entering the sample into the furnace. The flow is restored
as soon as any button on the PC-keyboard is pressed. For example taking a sample weight.
If an organic sample is entered into the furnace without oxygen flow and the flow is
subsequently restored, it may lead to a very intensive ignition. Therefore, the gas saving feature
for the resistance furnace operation should
only be installed after the customer’s consent to
take full responsibility for entering a sample into the furnace only after the flow is restored.
The modifications are shown in the following schematics. For full
Details, see chapter „
Wiring diagrams
“.
1.
Install a solenoid valve No. 01160 in the analyzer’s cabinet.
Fig. 100: Gas saving for resistance furnace operation
2.
Connect the valve in the tube leading to the resistance furnace inlet (close one port of the
T-piece of the valve, like already done for V3).
Fig. 101: Gas saving for resistance furnace operation
3.
Connect a cable to pin 49 of the 50-pin plug to the board HF-42.
V13
0
1 . 5
2 . 5
2
1
0 . 5
b a r
1 , 5 b a r / 2 2 p s i
Pressure switch
230VAC
regulator
Pressure
2-4 bar
30-60 psi
Oxygen inlet
PR1
2
0
0
l/
h
01160
3-21-1
Schematic 1
Gas saving for
resistance furnace
operation
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...