Faults
12
Pos : 3. 1 / 0025 Ü bersc hrift en/ELTR A 1. Ü berschrif ten/ 1 F aults Eltra @ 8\ mod_1396880810332_9. doc x @ 60058 @ 1 @ 1
3
Faults
Pos : 3. 2 / 0025 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 C hec k base lines ELTRA @ 6\ mod_1368452266990_9.doc x @ 40029 @ 2 @ 1
3.1 Check base lines
Pos : 3. 3 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 2000_s er vic e/05F aul ts_C S-2000/0505 M odul Chec k bas e lines @ 9\ mod_1425539018921_9.doc x @ 74093 @ @ 1
When this message appears on the PC screen, at least one of the base lines is either over
+9.5Volts or below -9.5 Volts.
In order to read the base line voltages of all IR cells available, click on “base lines”
Fig. 1: Base line button
A window appears at the right upper end of the screen showing the base lines moving from the
left to the right (time coordinate). In the middle of the right end of this base lines window, there
is a small button. After clicking on it, a new window appears, displaying the base line voltages,
described as “Inputs” and their “Averages”.
The description “Inputs” is made from the software point of view. The outputs of the cells are
inputs for the microcontroller board and for the software.
It is advisable to watch the figures under “Inputs”. Do not observe the “Averages” because the
response is delayed when the base lines change. This is a disadvantage when adjusting the
base lines.
Fig. 2: : Activating Inputs and Averages Window
Fig. 3: Inputs and Averages Window
If any of the “Inputs” or “Averages” is over 9.5V, adjust a voltage of about 6 to 7V.
See chapter 3.2
“
Infrared base line adjustment
”.
The adjustment is preferably made when having carrier gas flow through the cells.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...