Service
110
Pos : 7. 12 /0025 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Fl ow s ensor r epl ac ement Eltr a @ 8 \ mod_1396941567775_9. doc x @ 60384 @ 2 @ 1
5.6 Flow sensor replacement
Pos : 7. 13 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 15 Ser vic e_C S-2000/1535 Modul Flow sens or replac ement @ 9\ mod_1425543539438_9. doc x @ 74778 @ @ 1
Fig. 94: Flow sensor(3-14)
–
Set the power switch to pos. 0.
–
Remove the left side panel of the analyzer’s cabinet.
–
Remove the tubes (10) and (11) at the bottom-side of the infrared module.
–
Remove the plastic tubes from the left hand side of the IR module.
–
Remove the two plugs from the left hand side of the IR module.
–
Remove the screw fixing the IR module in the analyzer. Take the infrared module out of
the analyzer.
–
Remove the two tubes (5) and (6) from the flow sensor (9).
CAUTION
They shouldn’t be interchanged when reassembling! In case of doubt see chapter
“
IR-paths cleaning and replacing
”
–
Remove the two screws (7) underneath the flow sensor.
–
Remove the old flow sensor (9) and install a new one.
–
Reassemble the infrared rack in reverse order.
–
Pay particular attention to the position of the tubes (5), (6), (10) and (11). In case of doubt
see chapter “
IR-paths cleaning and replacing
”
–
Adjus
t the flow sensor. See chapter “
Gas flow controller
” where is written:
“Set the flow rate with P1 of the HF 42, to about 180 l/h. Observe the flow rate on the lower flow
meter of the analyzer.
Wait until the value is stable.”
1
9
5
6
7
11
10
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...