Faults
15
Check by removing the filter with the dust filter housing. See chapter:
“
High carbon results are
erratic
”.
Don’t remove the filter from its housing.
Turn the filter housing upside down and try to blow air through the smaller whole (about 16mm
diameter) of the filter housing.
If you feel a remarkable restriction, replace the filter by a new one. With a new filter the air
should pass practically free without any restriction.
Blocked chemicals:
The grains of magnesium perchlorate as well as of sodium hydroxide have to move when
knocking on their glass tube. If they are cloggy, they are completely depleted and they have to
be replaced. When the glass tube is turned upside down, the grains should fall off. If a screw
driver is needed to stab the grains free, the chemicals had to be replaced much earlier.
Blocked paper filters:
The paper filters are small so that they are frequently forgotten or neglected, although they can
block the flow when they are not replaced for long.
Blocked lance hole over the crucible:
The hole in the ceramic heat shield over the crucible in the induction furnace can be restricted
or blocked.
Clean by using the brass brush as described in the operation manual. If this may not be enough
then clean the hole by using a piece of wire or a pipe cleaner.
Blocked furnace outlet:
For induction furnace, unscrew the cap-nut at the furnace outlet and remove the plastic tube.
Pass through the fitting a pipe cleaner. Open the furnace in order to be sure that the pipe
cleaner has passed thru the fitting reaching the furnace area.
Completely remove the plastic tube by unscrewing the left hand cap-nut on the dust filter
assembly. Clean this fitting the same way like before. If necessary clean the plastic tube as well.
In case of resistance furnace operation clean the furnace outlet as shown below.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...