Faults
56
4
Bypass valve V4
5
Oxygen valve V5
6
Flow regulating valve V6
7
Purge valve V3
8
----
9
Exhaust valve V9
10
Valve V10
11
Solenoid valve V11
12
Solenoid valve V12
13
Solenoid valve V13
14
Gas outlet
15
Lance valve V2
16
Pressure switch
17
Oxygen pressure gauge 1,5 bar
18
Restrictor
19
Restrictor
20
Carrier gas flow meter
21
IR purge flow meter
22
Moisture trap after the furnace
23
CO
2
/ H
2
O
– purification trap
24
Catalyst furnace
25
SO
3
- trap
26
H
2
O
– trap
27
Dust filter
28
IR-cell
29
Gas pump
30
Attenuator before the pump
31
Attenuator after the pump
32
Resistance furnace
33
Induction firnace
34
Furnace input flow meter
35
Furnace inlet-oxygen
36
Furnace outlet measuring gas
37
Flow sensor
Detailed explanation of the faults indicated in the flow diagram
1.
Blockage of the lower connection of the glass tube (22)
The bore hole of the lower connection of the glass tube (22) can get blocked when there
isn’t enough quartz wool at the bottom end of the glass tube (22) enabling fine particles of
the magnesium perchlorate to fall into the bore blocking it. It is therefore absolutely
important to fill the glass tubes exactly according to the Instructions. See Operation
Manual “
Filling the reagent tubes
”.
Holes of other reagent tube connectors need to be cleaned from inside and outside of the
analyzer, due to the fact that the passage is not straight, but vertical outside and
horizontal inside the analyzer.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Page 6: ...Notes on service instructions 6 ...
Page 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Page 107: ...Service 107 Fig 91 Tubes to squeeze ...
Page 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Page 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Page 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Page 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Page 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Page 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Page 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Page 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Page 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...