ECP2010
/
ECP2010H
32 / 68
Action
Description
INIT PROGRAM
In item
Menu/Control/Init program
, press
ENTER.
Parameter setting of initialization step 0. On main screen displays
I►,
total time does not count up and flashes, pump runs.
Program start
On display is shown change from
I►
to
►,
total time and step time
starts. The method of start is described above.
Intercepting of
running program
a) Pressing
PUMP
On display is shown change from
►
to
II
, total time stops and flashes,
step time stops, pump stops. Mobile phase concentration change stops.
After pressing key PUMP or D again, program restarts.
b) Pressing
D
On display is shown change from
►
to
I►,
total time stops and flashes,
step time stops, pump is running. Mobile phase concentration change
stops.
By pressing
PUMP
status changes to held,
I►
changes to
II
and follows
manipulation as in previous paragraph.
By pressing
D
status changes to running
, I►
changes to
►.
REINIT PROGRAM
It is possible to proceed program reinitialization during program running
in
Menu/Control/Reinit program
. Prior to reinitialization it is possible to
hold the program by pressing key
PUMP
or
D.