ECP2010
/
ECP2010H
30 / 68
F
LOW
D
IARAM FOR
M
ENU
I
TEMS
C
ONTROL A
P
ROGRAM
S
ETUP
Steps from Program Creating to Leaving Program Mode.
a) Creating program in
EDIT PROGRAM
).
X. …
program sequence number
, S…Step, xx…
total number of steps, EMPT
Y…
empty.
b) Saving program. Then the program is available in
VIEW PROGRAM
) and
LOAD PROGRAM
c) Uploading a program into unit memory using
LOAD PROGRAM
. Unit turns into status of
control by program, which is called program mode. Screen switches to menu item
INIT
PROGRAM
.
d) Program initialization
INIT PROGRAM
(see
). The unit sets default program
parameters = initialization phase. Mobile phase is stabilizing as well as flowrate and
pressure. In this phase is a sample injected into loop.
e) Program start
–
by turning injecting valve from position LOAD to position INJECT (sample
gets to column); pressing key D or external signal, set in menu item
Menu/IO
Config/Interface/Digital input 1(2)/Program start
.
f) Program termination
–
automatically according to setting in menu item
END STEP
(final
step) -
RUN
(program ends, pump is running),
STOP
(program ends, pump stops)
,
REPEAT
(program ends, pump runs, program reinitiates for restart)
–
or during running
probram by evoking menu item
Menu/Control/Reinit program
and subsequently
Menu/Control/Stop program.
Programming mode remains saved after turning unit off and on. After restart it is in status
STOP
. If it is necessary to exit the program mode, it can be proceeded by
EXIT PROGRAM
MODE
(see
6.3.4.4
).