ECP2010
/
ECP2010H
47 / 68
If there occurs some deviation from standard unit
behavior, on main screen displays
W
with a
number of warnings.
The access to descriptions and other behavior is
the same as for errors, see chapter above.
6.6.3. Power
In this menu item is displayed value of main supply voltage, nominal value is 28.0 V and it
should be within the range of 26
–
30 V.
6.6.4. Cooling
Display shows
COOL
. Shows rotation speed of
FAN1
and
FAN2 (rpm)
and their
voltage
(V)
. In pump version ECP2010 / ECP2010H is only one fan mounted, therefore for one fan
is shown
NOT USED
.
6.6.5. Interface
On screen is displayed
IFACE
. This menu item gives information about inputs/outputs of
IO INTERFACE
, start signal input from injection valve (
START)
.
S
TATUS OF
IO
I
NTERFACE
Item
Description
AIN
Voltage on analog input.
DIN1 a DIN 2
Status of digital inputs.
DOUT
Status of digital output switch.
STRT
Status of input signal
START.
6.6.6. Leakage State
On screen is displayed
LEAK
. Status of leakage sensor.
S
TATUS OF
L
EAKAGE
S
ENSOR
Item
Description
VSPL
Voltage of pump measuring sensor L (L
… low
current).
VSPH
Voltage of pump measuring sensor H
(H …
high current).
VRPL
Voltage of pump reference sensor L.
VRPH
Voltage of pump reference sensor H.
VSBL
Voltage of box measuring sensor L.
VSBH
Voltage of box measuring sensor H.
VRBL
Voltage of box reference sensor L.
VRBH
Voltage of box reference sensor H.
CNTD
Countdown to current status change L/H, H/L (Countdown).
HEAT
Sensor heating
–
YES = heating (current High); NO = cooling- (current Low).
SPLK
Pump leakage sensor YES/NO.
SBLK
Box leakage sensor YES/NO.