ECP2010
/
ECP2010H
22 / 68
Value/Symbol
Meaning
IDD
Abbreviation of actual pump mode.
Actual piston position is displayed after first finished pump rotation.
During flowrate higher than ~3 ml/min actual position cannot be displayed
and this dummy symbol is shown.
Pulsation compensation status:
OFF
–
turned off.
FIXED
–
according to predefined values.
LEARNING
–
learning in progress.
LEARNED
–
according to learned values.
LEARNED OFF
–
compensation is turned off.
LEARNED 0
–
according to learned values (from FW 5.x).
LEARNED 0
–
compensation is turned off or not learned (from FW 5.x).
LEARNED 1
–
according to learned values (from FW 5.x).
LEARNED 1
–
compensation is turned off or not learned (from FW 5.x).
/ / / /
Gradient valve status (A, B, C, D). Only in gradient mode. During very fast
valve switching (low percentage of component and higher flowrate) it may
not manage to display valve switching.
Mark X
–
all valves are closed.
Indication of leakage, at the same time may be displayed an error or
warning (according to setting).
One IO INTERFACE inputs is used for pump control.
E(1)
An error has occurred
„ERROR“ (
number of errors). By using key ENTER
on main screen, it is possible to enter screen
QUICK DIAGNOSTICS
and
see a description of all errors and warnings.
W(1)
„WARNING“ (
number of warnings). By using key ENTER on main screen, it
is possible to enter screen
QUICK DIAGNOSTICS
and see a description
of all errors and warnings.
4.12. Door Installation
In order to prevent damage of door hinges, the door is supplied apart. The door should be
slipped on the hinges in opened position, see pictures below.
Door mounting must be easy,
no strength needed
, otherwise the hinges may get damaged.
Caution:
Never manipulate with the unit, when the door is mounted.