ECP2010
/
ECP2010H
48 / 68
6.6.7. Pump
Displays status of pump mechanism. Units for pressure are stated in upper row.
S
TATUS OF
P
UMP
M
ECHANISM
Item
Description
PACT
Actual pressure.
PAVG
Middle pressure during one cycle.
PULS
Pressure pulsation (bandwidth
PMAX
–
PMIN
) per one cycle.
DEAD
Status of optical switch detecting position of top dead center of main (left)
piston
–
CLOSE/OPEN.
MAGV
Motor load mean value (% of maximum allowed load).
MMAX
Motor load peak value (% of maximum allowed load).
CAMR
Camshaft speed (rotations per minute).
PPOS
Piston actual position.
6.6.8. Solvents
On main screen is displayed
SOLV
. In this menu item is displayed actual volume of residual
mobile phase (solvents).
6.6.9. Pulsation
On main screen is displayed
PULS
. In this menu item is displayed status of pump pulsation.
Units for pressure are stated in upper row.
P
UMP
P
ULSATION
S
TATUS
Item
Description
PU01
Pulsation during last rotation.
PUAN
Average pulsation in
n
rotations,
n
is determined by firmware based on
actual flowrate.
PUBT
Best reached pulsation during learning process.
PULI
Pressure pulsation limit.
PAVG
Average pressure in one rotation.
LTIM
Learning time
–
time lapsed from learning start.
LFLO
Learning flow
–
flowrate during learning (constant).
PPOS
Actual piston position.
Note:
After start of compensation learning (see
) for items PUAN, PUBT displayed
---. Values of pulsation are displayed after
n
rotations. At this moment is learning
process started.
6.6.10. Gradient State
On main screen is displayed
GRAD
. In this menu item is displayed status of values for
controlling gradient valve.