Place the unit in a suitable location that satisfies the following conditions:
•
Horizontal place.
•
Keep at least 10 cm space behind the rear of the unit.
•
Keep away from equipment generating strong magnetic field.
•
Equipment is intended for use in regular laboratory environment only - see
Technical
parameters
–
operating environment conditions
.
4.1. Inlet Tubing Connection
For connecting pump in
isocratic
mode, use inlet tubing with filter (ABA16000) from supplied
accessories. For
gradient
mode connection is supplied inlet tubing (ABA19000), which is among
accessories of the box.
At first unscrew the white cap on pump inlet.
Caution:
The pump is filled with isopropanol after construction or service.
Isocratic connection of pump
inlet
Connect inlet tubing with filter.
Snap the tubing to the socket for
tubing and capillaries.
Set of inlet tubing with a lid and
filter (picture on the right).
Gradient connection of pump
inlet
Connect inlet tubing. Connect the
other end to the
box.
Tubing with moving compensation
is fixed into the upper socket for
tubing and capillaries.
Note:
The upper and side sockets in the unit front panel are determined for gripping
tubing and capillaries. FEP tubing with outer diameter
1/8“
are fixed by snapping
into the slot. Capillaries and tubing with outer diameter 1
/16“
are fixed using a
short FEP tubing with outer diameter 1
/8“ a
nd inner diameter 1/16
“,
threaded on
the capillary. Such a threaded tubing cannot be snapped in the slot, it should be
pushed in from the side.
4.2. Outlet Capillary Connection
To connect the pump to injector, use outlet capillary (ACA99300). For connecting pump to column,
use outlet capillary (ACA99400).