25 / 68
6.2.1. General
In these sub menus can be set pressure units, display brightness, sounds and pump mode.
6.2.1.1. Pressure Units
Pressure units can be set to
MPa
(megapascal),
PSI
(pound-force per sq. inch)
and
BAR
(bar). 1 MPa =10 bar = 145 psi. Default pressure unit is MPa.
Note:
The PSI unit pressure value are automatically rounded to tens (ie. 145 psi after
rounding is 150 psi).
6.2.1.2. Display Brightness
Display brightness can be set in 8 levels, where 8 is maximum brightness. Default value is 3.
6.2.1.3. Sound
Sound signals of the unit can be set as follows:
S
OUND
S
ETTING
Setting
Description
KEYBOARD
At each touch of button will be generated short beep when
the key does not have any significance at the moment, mid-
length beep for value change and moving in menu and longer
beep for confirmation of value changes.
ERRORS
At error occurrence will be generated sound signal of the
error. Three consecutive longer beeps in case of an error and
in case of a fatal error with repetition every 5 s. Repetition
can be interrupted by pressing any key.
WARNINGS
At warning occurrence will be generated sound signal of
warning. Three consecutive shorter beeps.
OPERATIONS
During operation will be generated longer beep and for long
operations may be this beep generated at the beginning and
at the end of an operation.
In default setting are all sounds set on.
6.2.2. Purge
Here can be set possibilities of pump purge.
6.2.2.1. Purge Flow
Purge flow can be set in range
1.00 - 12.00 ml/min
. Default value is 12 ml/min.
6.2.2.2. Pressure Limit
Purge pressure limit can be set in range
0.1 - 2.0 MPa
(10 - 290 psi)
. Default value is 1 MPa
(150 psi).
6.2.2.3. Purge Time
Purge time can be set in range
0 - 10 min
. Value 0 means that purge is not time limited.
Default value is 0 min.