4-46
Communication Interfaces
HPG 13150 400 V (A)
Dressler
®
HF-Technik GmbH
Prerelease version
17
frequency
tuning
parameters
Sets the parameters for frequency
tuning.
Send three data bytes, least significant
byte first.
• Bytes1 and 2 specify the maximum
frequency in kHz (this value is
ignored when byte 3 is set to 0).
• Byte 3 selects the control mode for
frequency shift tuning (FST):
4
0 = No FST (frequency as defined
with command
18
)
4
1 = Integral, phase positive
4
2 = Integral, phase negative
4
3 = Proportional, phase positive
4
4 = Proportional, phase negative
If FST is active (modes 1 through 4), the
frequency set with command
18
is used
as the minimum frequency limit. Ensure
that this frequency is less than the
maximum frequency before command
17
is used.
Note:
This command is not available on
generators with fixed frequency.
Read back with
187
.
3
1
(CSR only)
18
AE Bus and
Ethernet
version
RF
frequency
(see next
row for
Profibus
version of
command)
Sets the RF frequency.
Send four data bytes representing the RF
frequency in 100 Hertz steps (for
example, 10000 = 1 MHz).
Note:
This command is only available on
generators with variable frequency.
Read back with command
178
.
4
1
(CSR only)
Table 4-13. Host port commands (Continued)
Command
Description
Number of
Host Data
Bytes
Number of
Response
Data Bytes
Summary of Contents for 13150 400 V
Page 2: ......
Page 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Page 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...