5-32
Installation, Setup, and Operation
HPG 13150 400 V (A)
Dressler
®
HF-Technik GmbH
Prerelease version
To Read the Tube Lifetime Parameters:
1. Press the
Program
soft key.
2. Press
Next
until you see
Tube Lifetime
on the display. This screen shows
the following parameters:
4
Lifetime counter in hours
4
Lifetime counter state (displays OK, warning, or life end)
3. To enable or disable the tube lifetime counter, press the
Mode
soft key.
(Disabling the tube lifetime counter turns off the tube lifetime warning.
Even if the counter is disabled, the unit still keeps an internal record of the
lifetime.)
4
To enable the tube lifetime counter, press the
Counter
soft key.
4
To disable the counter, press the
Off
soft key.
4. To exit, press
Escape
.
CHANGING THE PULSE INPUT CONFIGURATION SETTINGS
Before changing the pulse input configuration settings, you must first turn on
the pulsing function. For more information, see
“To Select Internal or External
Pulsing Using the HPG Generator Front Panel:” on page 5-25
The configuration settings allow you to program the “polarity” of the HPG
generator's pulse input function.
To Set the Pulsing Input Parameters Using the HPG Generator Front Panel:
1. Press the
Program
soft key.
2. Press
Next
until you see
Pulse Input Configuration
on the display.
3. Press
Change
.
4. Select the desired “polarity” of the pulse input:
a. If the pulse function is set to external pulsing, choose either:
4
Hi=RFon
Low=RFof
f
This setting enables standard external pulsing.
4
Hi=RFoff
Low=RF on
This setting enables inverse external pulsing.
b. If the pulse function is set to internal pulsing, choose from:
4
Hi=CW
Low=Pulse
Summary of Contents for 13150 400 V
Page 2: ......
Page 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Page 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...